• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WESTMED LLC ADULT SOFT TIP CANNULA WITH 7' 85 SHORE GREEN TUBING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WESTMED LLC ADULT SOFT TIP CANNULA WITH 7' 85 SHORE GREEN TUBING Back to Search Results
Model Number 0556
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 09/26/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The leaks or loose pipes at the y connector would interrupt the patient therapy, and the patient could become hypoxia.
 
Event or Problem Description
Cannulas are leaking. .
 
Event or Problem Description
Cannulas are leaking. .
 
Additional Manufacturer Narrative
The leaks or loose pipes at the y connector would interrupt the patient therapy, and the patient could become hypoxia no images or returned inventory provided, complaint is unconfirmed.The issue is stated as the cannula is leaking.There was nothing in inventory to review as well as nothing out of the ordinary for the dhr's.In the last 24 months, there have been 2 related complaints.Root cause cannot be determined at this time.However, likely root cause is little or no bonding agent applied to bonding location during assembly.Risk assessment: for patient risk, rma-20017f, step r87 issue is stated as "low oxygen delivery", potential hazard being "loss/delay in oxygen therapy leading to hypoxia", causation "cannula tubing deteriorates resulting in leaks prior to first use", s=8, o=1, rpn=8 (less than 25, therefore risk is acceptable).For regulatory risk the lack of directly related complaints indicates a low-risk.
 
Additional Manufacturer Narrative
The leaks or loose pipes at the y connector would interrupt the patient therapy, and the patient could become hypoxia.No images or returned inventory provided; complaint is unconfirmed.The issue is stated as the cannula is leaking.There was nothing in inventory to review as well as nothing out of the ordinary for the dhr's.In the last 24 months, there have been (b)(4) related complaints.Root cause cannot be determined at this time.However, likely root cause is little or no bonding agent applied to bonding location during assembly.Risk assessment: for patient risk, rma-20017f, step r87 issue is stated as "low oxygen delivery", potential hazard being "loss/delay in oxygen therapy leading to hypoxia", causation "cannula tubing deteriorates resulting in leaks prior to first use", s=8, o=1, rpn=8 (less than 25, therefore risk is acceptable).For regulatory risk the lack of directly related complaints indicates a low-risk.Udi related data quality updates only.
 
Event or Problem Description
Cannulas are leaking. .
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADULT SOFT TIP CANNULA WITH 7' 85 SHORE GREEN TUBING
Common Device Name
ADULT SOFT TIP CANNULA WITH 7' 85 SHORE GREEN TUBING
Manufacturer (Section D)
WESTMED LLC
5580 s nogales highway
tuscon AZ 85706
Manufacturer (Section G)
WESTMED LLC
5580 s nogales highway
tuscon AZ 85706
Manufacturer Contact
melissa brickley
2710 northridge dr nw suite a
grand rapids, MI 49544
6162598415
MDR Report Key18196862
Report Number2028807-2023-00032
Device Sequence Number6554120
Product Code CAT
UDI-Device Identifier00709078002960
UDI-Public(01)00709078002960(10)032622J03,(01)00709078002960(10)032022J03
Combination Product (Y/N)N
Initial Reporter CountryFR
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse Practitioner
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0556
Device Catalogue Number0556
Device Lot Number032622J03, 032022J03
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/25/2023
Supplement Date Received by Manufacturer10/25/2023
10/25/2023
Initial Report FDA Received Date11/22/2023
Supplement Report FDA Received Date01/30/2024
07/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;