| Catalog Number |
320-15-04 |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Aseptic loosening (4883)
|
| Date of Event |
06/28/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(h3) pending evaluation (d10) concomitant device(s): (b)(6): 300-30-06 - equinoxe preserve stem 6mm.(b)(6): 320-10-00 - equinoxe reverse tray adapter plate tray +0.(b)(6): 320-15-04 - rs glenoid plate r post aug, 8 deg, right.(b)(6): 320-46-03 - equinoxe reverse 46mm humeral liner +2.5.
|
| |
|
Event or Problem Description
|
|
As reported by the equinoxe shoulder study, the patient had an initial tsa on (b)(6) 2020.The patient presented on 28-jun-2023 with aseptic glenoid loosening - breakage of components leading to glenoid loosening.The case report form indicates that this event is possibly related to device and/or to procedure.Outcome is continuing with instructions to watch & wait, stoke patient waiting medical clearance.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The reason for the revision reported cannot be confirmed from the information provided but may be the result of aseptic glenoid loosening and screw fracture.However, the loosening and fracture cannot be confirmed and potential contributions of user or patient-related considerations could not be assessed as the component was not returned for evaluation and no images, radiographs, or relevant clinical information were provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Search Alerts/Recalls
|