• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SET MENISCUS MENDER II DISPOSABLE; PASSER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SET MENISCUS MENDER II DISPOSABLE; PASSER Back to Search Results
Catalog Number 7209485
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/03/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during an arthroscopy, the set meniscus mender ii was broken when the package was opened.The procedure was successfully completed with non-significant surgical delay using a smith and nephew back up device.No further complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Additional Manufacturer Narrative
The reported device was received for evaluation.A visual inspection of the returned device found that it is not in its original packaging.The meniscus mender set was returned with one of the suture retriever shafts fractured away from the hub.This set of devices was manufactured before the hub was reinforced to alleviated this issue.A complaint history review found similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the device drawings found that a certificate of conformance to material and processing specifications must be provided.Weld strength requirements are specified.The root cause was associated with device design.A design change was implemented as a result of a corrective action initiated to address the reported issue.A stress relief collar is now attached at the connection between the shaft and the hub of the suture capture loop.Futher corrective actions are not recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SET MENISCUS MENDER II DISPOSABLE
Common Device Name
PASSER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18198490
Report Number1219602-2023-02278
Device Sequence Number7467483
Product Code HWQ
UDI-Device Identifier03596010468482
UDI-Public03596010468482
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K885311
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/16/2026
Device Catalogue Number7209485
Device Lot Number2065518
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/03/2023
Supplement Date Received by Manufacturer03/05/2024
Initial Report FDA Received Date11/23/2023
Supplement Report FDA Received Date03/06/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age26 YR
Patient SexFemale
Patient Weight45 KG
-
-