BIOMET 3I OSSEOTITE TAPERED IMPLANT 4 X 8.5MM; DENTAL IMPLANT Back to Search Results
Catalog Number NT485
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/25/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the mount was damaged.While tryng to remove the mount after implant placement, it was impossible to unscrew it and implant had to be removed.Tooth site 37.Procedure was completed with other implant.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).A4: patient weight unknown / not provided.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).The following fields have been updated: b4: date of this report.B5: describe event or problem.D4: additional device information.D9: device availability.G3: date received by manufacturer.G6: type of report.H1: type of reportable event.H2: follow up type.H3: device evaluated by manufacturer.H4: device manufacturer date.H6: adverse event problem.H10: additional narrative.Zimvie received one (1) nt485, (osseotite® tapered implant 4 x 8.5mm) for evaluation.Visual evaluation of the as returned device identified the implant and mount with signs of use.Mount did not disengage from the implant during a functional / physical test.Dhr review was completed for the subject lot number 2023060727.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2023060727 for similar events and no other complaint was identified.Review completed utilizing keywords: does not disengage/release.Based on the investigation and risk management file review, the most likely root causes determined from the investigation are excessive torque applied - customer error.Therefore, based on the available information, and functional test a device malfunction did occur.The reported event was confirmed with all the available information.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSSEOTITE TAPERED IMPLANT 4 X 8.5MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key18203972
Report Number0001038806-2023-02254
Device Sequence Number7466879
Product Code DZE
UDI-Device Identifier00844868012054
UDI-Public(01)00844868012054(17)280626(10)2023060727
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K063286
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 05/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberNT485
Device Lot Number2023060727
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/05/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date11/24/2023
Supplement Report FDA Received Date05/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/25/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age53 YR
Patient SexFemale
Patient EthnicityNon Hispanic
-
-