| Model Number |
PSO-PTT |
| Medical Device Problem Code |
Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/30/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The machine showed e005 after the catheter was implanted for 12 hours.
|
| |
|
Additional Manufacturer Narrative
|
|
The catheter in return is not compliant, error e005 is confirmed by the breakage of the temperature micro-cable in the dongle shell.Excessive and irreversible abnormal traction while using the catheter as it has been operating for 12 hours.User error confirmed.
|
| |
|
Event or Problem Description
|
|
The machine showed e005 after the catheter was implanted for 12 hours.
|
| |
|