| Model Number |
PSO-VT |
| Medical Device Problem Code |
Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The machine showed e001 after the catheter was implanted for 24hours.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis: soiled catheter with orange-red deposits (blood) in the drainage holes and in a large part of the central channel.A suture still attached to the silicone tube.A micro-thread (connection between sensor and micro-cable) under the silicone capsule appears broken at the edge of the capsule.Conclusion: the return catheter is not compliant, with a break under the silicone of one of the micro-wires connecting the micro-cable to one of the sensor pads causing error e001.
|
| |
|
Event or Problem Description
|
|
The machine showed e001 after the catheter was implanted for 24 hours.
|
| |
|