| Model Number |
799-EMVP-04 |
| Medical Device Problem Code |
Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/05/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that while attempting to treat a 62-year-old male patient, the device inappropriately shut down.Complainant indicated that the clinician bag ventilated to continue treating the patient.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation; the customer's report was duplicated and attributed to loose hardware from the dovetail bracket shorted with the power interface module (pim) board.The pim board was replaced to remedy the report.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental medwatch report updates information based upon your request which differs from the initial medwatch report filed.Please reference updated section d1 brand name, and section d4 catalog number that does not apply and should be removed.
|
| |
|