Catalog Number 02-022-35-3512
Medical Device Problem Codes Naturally Worn (2988); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/20/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
D10: concomitants: 6025476 201-78-89 - 3" drill bit, mod.Hex 2 pack.6132385 201-78-82 - collar fixation pin 2pk 40mm, quick release.6219095 200-02-29 - three peg patella 29mm.6378910 02-022-45-3525 - truliant tib fit tray cem sz 3.5f / 2.5t.6548841 02-020-11-0335 - truliant ps cem fem ps cem right sz 3.5.6631806 201-78-25 - small headed sharp pin, 1 1/8" 4 pack.Pending investigation.
 
Event or Problem Description
As reported via legal documentation the patient had a right knee replacement on (b)(6)2020.Approximately 2 years and 10 months after the initial procedure the patient had a right knee revision on (b)(6) 2023.There is no other patient demographic or medical history available.There is no information on the surgical procedure or patient outcome.There is no device return.There are no photos or other images of the device provided.No additional information is available.The patient has filed a complaint in a coordinated action in (b)(6) county with master case no.(b)(4).The consolidated long form complaint that applies to cases filed in this coordinated action alleges that patients filing suits in this coordinated action were required ¿to undergo revision surgeries due to severe, pain, swelling, and instability¿ due to ¿wear of the polyethylene components and resulting component loosening and/or other failure failures causing serious complications including tissue damage, osteolysis, permanent bone loss, and other injuries.¿ because the patient has filed a short-form complaint in this coordinated action, the patient appears to allege that the patient was injured as a result of wear of an exactech polyethylene device.
 
Additional Manufacturer Narrative
H6: corrected the following: medical device problem code, component code, type of investigation, investigation findings, investigation conclusions.Manufacturer narrative updated: the reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear or due to inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, and relevant clinical information were not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRULIANT TIB IMP PS INSERT SZ 3.5 12MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th ct.
gainesville FL 32653
Manufacturer Contact
geoff gannon
MDR Report Key18209868
Report Number1038671-2023-02860
Device Sequence Number15069058
Product Code JWH
UDI-Device Identifier10885862304377
UDI-Public10885862304377
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K171045
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 10/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/18/2025
Device Catalogue Number02-022-35-3512
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/05/2023
Supplement Date Received by Manufacturer10/14/2024
Initial Report FDA Received Date11/27/2023
Supplement Report FDA Received Date10/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/21/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0023-2022
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-