• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LGN POROUS CR FEM SZ 4L; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LGN POROUS CR FEM SZ 4L; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 71423244
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Loss of Range of Motion (2032)
Date of Event 10/11/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference case-(b)(4).
 
Event or Problem Description
It was reported that, after a tka surgery performed (b)(6) 2023, the patient experienced an unknown adverse event.The issue was solved with manipulation under anesthesia with intraarticular steroid and local anesthetic injection.The patient current health status is unknown.Further information is not available.
 
Additional Manufacturer Narrative
Section h3, h6: given the nature of the alleged incident, the devices could not be returned for evaluation.The clinical/medical investigation concluded that, the electronic clinical report form was provided and confirms the compliant however, the do not aid in determining the root cause.As of the date of this medical investigation, the requested clinical documentation has not been provided; therefore, there were no clinical factors found which would have contributed to the event.The instructions for use for the knee system does warn, ¿post-operative therapy should be structed to prevent excessive loading of the operative knee and to encourage bone healing.¿ according to the report, the issue was solved with manipulation under anesthesia with intraarticular steroid and local anesthetic injection.The impact to the patient beyond the unknown adverse event, the manipulation under anesthesia and the medications cannot be determined since the current health status of the patient is unknown.A review of the production orders did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed a similar event for the part numbers for baseplate and the insert over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.This similar event reported an adverse event but no sufficient information about the failure mode was provided, so these events could not be associated with the reported event.For the femoral component a review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the instructions for use documents for knee systems revealed that it a provides complete guidelines of indications, contraindications, warnings and precautions and possible adverse effects that may occur preoperative, during surgery or post operative.As no details of the alleged fault, malfunction or injury were provided a risk management files and prior actions review could not be performed.At this time, we have no reason to suspect that the product failed to meet any specifications at the time of manufacture.No details of the alleged fault, malfunction or injury were provided, therefore no factors that can contribute can be delineated.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN POROUS CR FEM SZ 4L
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18209916
Report Number1020279-2023-02325
Device Sequence Number15014556
Product Code MBH
UDI-Device Identifier00885556030936
UDI-Public00885556030936
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K073325
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number71423244
Device Lot Number20MHA0035
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/31/2023
Supplement Date Received by Manufacturer01/05/2024
Initial Report FDA Received Date11/27/2023
Supplement Report FDA Received Date01/08/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/03/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age55 YR
Patient SexFemale
Patient Weight89 KG
-
-