| Catalog Number |
60SP030 |
| Medical Device Problem Codes |
Break (1069); Fracture (1260); Material Separation (1562)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
11/14/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3: other; device not returned to manufacturer.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that "the collar cracked while the tube was changed¿.The event occurred at the patient¿s home during patient use.Adverse patient effects are unknown.The customer stated that no further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
D9 device returned to mfg: 1/3/2024.One device was received for investigation.The reported complaint was confirmed during visual inspection, which verified the device was broken at the flange.The investigation attributed this condition to the device having come into contact with a sharp object during reprocessing, or another issue related to handling.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.Complaint data related to the reported issue will continue to be monitored.
|
| |
|
Additional Manufacturer Narrative
|
|
D4 was corrected.**udi related data quality updates only**.
|
| |
|