• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI ARCTIC FRONT ADVANCE PRO¿ CARDIAC CRYOABLATION CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI ARCTIC FRONT ADVANCE PRO¿ CARDIAC CRYOABLATION CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Model Number AFAPRO28
Medical Device Problem Codes Compatibility Problem (2960); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/26/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product event summary: the afapro28 balloon catheter with lot number 06490 was returned and analyzed.Visual inspection of the balloon, shaft, and handle segments was performed and no anomalies were identified.The catheter smart chip data was downloaded and reviewed.The data indicated that the balloon catheter was never used.The balloon catheter was recognized and passed the electrical integrity verifications and performance test.The balloon catheter completed the inflation, ablation, and thawing phases with no console system notices generated.No performance issues were identified, all pressure and flow values were in range, and the temperature curve had no oscillations or overshoots.Functional testing was performed and the mapping catheter was unable to be inserted into the balloon catheter.Pressure testing and inspection of the sub-components of the catheter, handle, and shaft segments was performed and a kink on the guide wire lumen was observed inside of the catheter handle (between the y-block and service loop) after dissection into the balloon catheter.No leakage at the kink location was identified.In conclusion, the reported compatibility issue between the balloon catheter and mapping catheter was confirmed.The balloon catheter failed the returned product inspection due to a kink/ twist on the guide wire lumen.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during a cryo ablation procedure, the mapping catheter would not advance into the balloon catheter.The balloon catheter was replaced to resolve the issue. the case was completed with cryo. no patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTIC FRONT ADVANCE PRO¿ CARDIAC CRYOABLATION CATHETER
Common Device Name
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key18221223
Report Number2649622-2023-33732
Device Sequence Number6622974
Product Code OAE
UDI-Device Identifier00763000341503
UDI-Public00763000341503
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P100010
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/14/2023
Device Model NumberAFAPRO28
Device Catalogue NumberAFAPRO28
Device Lot Number06490
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/31/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/02/2023
Initial Report FDA Received Date11/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/14/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age62 YR
Patient SexMale
-
-