MEDTRONIC MEXICO S. DE R.L. DE CV BROCKENBROUGH¿ NEEDLE; TROCAR Back to Search Results
Model Number EP003994S
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/17/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a transeptal procedure, the needle could not advance through the sheath.The sheath was replaced, but the issue persisted.Saline was attempted to be flushed through the sheath, but remained within the sheath, unable to advance through the tip.The sheath and needle were replaced which resolved the issue.The case was completed. no patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Continuation of d10: product id 990061-070 product type sheath; product id 990061-070 product type sheath product event summary: the ep003994s transseptal needle with lot number (b)(6).Was returned and analyzed.The transseptal needle was visually inspected and functionally tested and no visible blood was found on the needle.The transseptal needle was intact with no apparent issues, no kink or bend was noted along the needle, however the needle was unable to enter into the returned dilator sheath.Upon evaluation of the returned sheath dilators, it was found that there was skived material and skive marks inside of both sheath dilators when checked with a borescope.In conclusion, the product analysis of the transseptal needle confirmed the transseptal compatibility issue.The transseptal needle passed the returned product inspection as per specification.However, when the sheath dilators were analyzed, it was found that there was skived material and skive marks inside of both dilators.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Udi related data quality updates only.Corrected: d4 (provide complete udi).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BROCKENBROUGH¿ NEEDLE
Common Device Name
TROCAR
Manufacturer (Section D)
MEDTRONIC MEXICO S. DE R.L. DE CV
av. paseo del cucapah #10510
tijuana,bc 22210
MX  22210
Manufacturer (Section G)
MEDTRONIC MEXICO S. DE R.L. DE CV
av. paseo del cucapah #10510
tijuana,bc 22210
MX   22210
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key18221259
Report Number9617601-2023-00238
Device Sequence Number12163096
Product Code DRC
UDI-Device Identifier00613994663146
UDI-Public(01)00613994663146(17)241207(1
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
PRE-AMEND
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/07/2024
Device Model NumberEP003994S
Device Catalogue NumberEP003994S
Device Lot Number225304499
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/13/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/01/2023
Supplement Date Received by Manufacturer11/01/2023
Initial Report FDA Received Date11/28/2023
Supplement Report FDA Received Date07/24/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/08/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10....
Patient SexUnknown
-
-