| Model Number |
CD1231-40Q |
| Medical Device Problem Codes |
Premature Discharge of Battery (1057); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/08/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported during in clinic follow up that the implantable cardioverter defibrillator exhibited premature battery depletion.The device was explanted and successfully replaced.The patient was stable and asymptomatic.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported field event of premature battery depletion was not confirmed.Final analysis found the device was at elective replacement indicator (eri) upon receipt.Longevity analysis found the battery depletion to be normal.The device functioned normally during analysis.No anomalies were found.The event of longevity overestimation was confirmed.The overestimation was determined to be due to an estimation inaccuracy in the merlin programmer remaining longevity calculation.
|
| |
|
Event or Problem Description
|
|
Related manufacturer reference number: 2017865-2024-32769.
|
| |
|