• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) FORTIFY VR, U1.6 SJ4 US; PULSE GENERATOR, PERMANENT, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) FORTIFY VR, U1.6 SJ4 US; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number CD1231-40Q
Medical Device Problem Codes Premature Discharge of Battery (1057); No Apparent Adverse Event (3189)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/08/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported during in clinic follow up that the implantable cardioverter defibrillator exhibited premature battery depletion.The device was explanted and successfully replaced.The patient was stable and asymptomatic.
 
Additional Manufacturer Narrative
The reported field event of premature battery depletion was not confirmed.Final analysis found the device was at elective replacement indicator (eri) upon receipt.Longevity analysis found the battery depletion to be normal.The device functioned normally during analysis.No anomalies were found.The event of longevity overestimation was confirmed.The overestimation was determined to be due to an estimation inaccuracy in the merlin programmer remaining longevity calculation.
 
Event or Problem Description
Related manufacturer reference number: 2017865-2024-32769.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORTIFY VR, U1.6 SJ4 US
Common Device Name
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18221690
Report Number2017865-2023-93350
Device Sequence Number5671536
Product Code NVZ
UDI-Device Identifier05414734504348
UDI-Public05414734504348
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P910023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician Assistant
Type of Report Initial,Followup
Report Date (Section B) 02/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2013
Device Model NumberCD1231-40Q
Device Lot Number3515235
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/08/2023
Supplement Date Received by Manufacturer02/15/2024
Initial Report FDA Received Date11/28/2023
Supplement Report FDA Received Date02/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/29/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DURATA.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age48 YR
Patient SexFemale
-
-