| Model Number |
39200 |
| Medical Device Problem Codes |
Material Deformation (2976); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/12/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 09nov2023.It was reported that the shaft was kinked and damaged.After unpacking an 8x60x120 epic vascular stent, it was observed that the shaft catheter was kinked and damaged and could not be implanted.The procedure was then completed with a different device.There were no patient complications reported, and the patient was stable post-procedure.However, returned device analysis revealed that stent was partially deployed.
|
| |
|
Additional Manufacturer Narrative
|
|
E1: initial reporter phone: (b)(6).Device evaluated by mfr.: epic vascular 8x60x120, was received for analysis.The device was received with the safety clip removed and the stent partially deployed on the delivery system.No issues were noted with the stent.A visual examination found no kinks or damage to the outer sheath of the device.A visual inspection found the inner sheath of the device to be kinked at approximately 3mm proximal of the tip bond.This type of damage is consistent with excessive force being applied to the device.A visual and tactile examination identified no issues with the tip of the device.No damage or issues were noted with the handle of the device.The safety lock had been removed from the rack.
|
| |
|