• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LDR MEDICAL ROI-C LONG ANCHORING PLATE; ROI-C TITANIUM-COATED IMPLANT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LDR MEDICAL ROI-C LONG ANCHORING PLATE; ROI-C TITANIUM-COATED IMPLANT SYSTEM Back to Search Results
Catalog Number MC1006T
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/06/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
 
Event or Problem Description
It was reported that the roi-c implant was placed in the patient and that upon malleting the first long plate, it became quite difficult to advance the last 2mm.The surgeon had to mallet quite hard to finish placing the long plate in the correct position.The second long plate was also more difficult than usual.The surgeon did not use the awls as he did not believe the bone was hard enough to merit them.Upon removal of the inserter handle, it was noted that a piece of the peek implant had chipped off.The surgeon tugged on the plates to ensure they were secure without the chipped piece and the implant was left in place.There was a 10-minute delay without impact on the patient.This is report two of two for this event.
 
Additional Manufacturer Narrative
Additional information in h6: component, investigation type, findings, and conclusions.Inspection the device was not returned and no photos were provided, so an evaluation is unable to be performed.Dhr review the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left highridge¿s control.Potential root cause a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to the awl not being used during surgery.Device usage this device is used for treatment.A follow-up report will be sent if additional information is obtained that adds value to the relevant content of this report.Reference report 3004788213-2023-00137.
 
Event or Problem Description
It was reported that the roi-c implant was placed in the patient and that upon malleting the first long plate, it became quite difficult to advance the last 2mm.The surgeon had to mallet quite hard to finish placing the long plate in the correct position.The second long plate was also more difficult than usual.The surgeon did not use the awls as he did not believe the bone was hard enough to merit them.Upon removal of the inserter handle, it was noted that a piece of the peek implant had chipped off.The surgeon tugged on the plates to ensure they were secure without the chipped piece and the implant was left in place.There was a 10-minute delay without impact on the patient.This is report two of two for this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROI-C LONG ANCHORING PLATE
Common Device Name
ROI-C TITANIUM-COATED IMPLANT SYSTEM
Manufacturer (Section D)
LDR MEDICAL
5 rue de berlin
quartier europe de l'ouest
sainte-savine 10300
FR  10300
Manufacturer (Section G)
LDR MEDICAL
5 rue de berlin
quartier europe de l'ouest
sainte-savine 10300
FR   10300
Manufacturer Contact
sabrina abla
10225 westmoor dr.
westminster, CO 80021
7206965158
MDR Report Key18222549
Report Number3004788213-2023-00138
Device Sequence Number6721386
Product Code OVE
UDI-Device Identifier03662663016988
UDI-Public(01)03662663016988(17)270901(10)1003144
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K151934
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberMC1006T
Device Lot Number1003144
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date11/28/2023
Supplement Report FDA Received Date06/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age51 YR
Patient SexMale
Patient EthnicityNon Hispanic
-
-