ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. 107" (272 CM) APPX 8.7 ML, TRANSFER SET W/DUAL CHECK VALVE, MICROCLAVE®, LUER LO; STOPCOCK, I.V. SET
|
Back to Search Results |
|
| Catalog Number |
B33895 |
| Medical Device Problem Code |
Precipitate in Device or Device Ingredient (1478)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The event involved a 107" (272 cm) transfer set w/dual check valve, microclave®, luer lock.It was reported started to see some issues when incompatible medications are administered through the line.They administer one medication, do a good flush and then the next one, which may be incompatible.That is when they were noticing precipitate and just had a line clot off.The event occurred while the product was in use with a patient, after a calcium infusion and flush, during the antibiotic infusion.Ms.Shafland stated that there was human harm reported to be associated with the event, it clotted off the line but they were able to save it.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned for evaluation; however, testing has not yet been completed.
|
| |
|
Additional Manufacturer Narrative
|
|
The possible lot number was reviewed, no nonconformities were identified that may have contributed to the reported complaint.The b33895 transfer set and connected componentry were evaluated and tested and there was no evidence of leakage or precipitate observed.The complaint was unable to be confirmed.
|
| |
|