DEPUY SYNTHES PRODUCTS LLC 7.5CM ANGLE ATTACHMENT; DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC)
|
Back to Search Results |
|
| Catalog Number |
QD8-S |
| Medical Device Problem Code |
Excessive Heating (4030)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthese, or its employees that the report constitutes an admission that the product, depuy synthese, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed, and it was determined that the reported condition that the device produced heat, identified during service and repair, was confirmed.The assignable root cause was determined to be due to component failure from wear.Udi: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported by switzerland that during service and evaluation, it was determined that the attachment device produced heat and vibration.It was further determined that the device failed pretest for vibration and temperature.It was noted in the service order that during pre-surgery, it was discovered that the device was not working properly it was not reported if there were any delays in the surgical procedure or if a spare device was available for use.There was no patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2023.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|