EXACTECH, INC. NV GXL LNR, LIPPED, 28MM ID, GROUP 1 CUPS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
|
Back to Search Results |
|
| Catalog Number |
132-28-51 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Synovitis (2094); Osteolysis (2377)
|
| Date of Event |
08/30/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Pending investigation.
|
| |
|
Event or Problem Description
|
|
As reported via legal documentation the patient had a left hip replacement on (b)(6) 2009.Approximately 13 years and 8 months after the initial procedure the patient had a left hip revision on (b)(6) 2023.There is no other patient demographic or medical history available.There is no device return.There are no photos or other images of the device provided.No additional information is available.Revision op report on (b)(6) 2023.Diagnosis: left total hip osteolysis, polywear.The acetabular liner was found to be worn with evidence of delamination.The acetabular component was found to be well-fixed with surrounding osteolysis, especially posteriorly.The patient was transferred to the bed and then recovery room in stable condition.
|
| |
|
Additional Manufacturer Narrative
|
|
H10.Updated/additional information ¿ the most likely cause for the revision reported due to prosthesis wear is a combination of risk factors including, use error, implant positioning, implant size selection, and patient factors may have also been a contributing factor to the early prosthesis wear.However, this cannot be not confirmed with the information provided; devices were not returned.There is no other information available.These devices are used for treatment not diagnosis.
|
| |
|
Search Alerts/Recalls
|
|
|