| Catalog Number |
PRO2043 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is not expected to be returned for evaluation and review.However, the complaint investigation is not complete at this time.A supplemental and final report will be filed following the completion of the complaint investigation.We will continue to monitor for trends through the complaint system to assure patient safety.
|
| |
|
Event or Problem Description
|
|
A sales representative reported on behalf of a customer that the device, pro2043, powerpro sagittal saw attachment, was being used during a laterjet procedure on 01nov23 and "the powerpro sagittal saw attachment was being used with a 90 degree saw blade.When applying force the cut, the blade came loose and shot out of the hub nearly hitting a staff member¿.The procedure was completed using a pro8250sb, and there was a delay of ten minutes.There was no report of injury to the patient or user.A good faith effort was made to obtain additional information; no response was received.This report is being raised due to the reported malfunction with potential for injury upon reoccurrence.
|
| |
|
Event or Problem Description
|
|
A sales representative reported on behalf of a customer that the device, pro2043, powerpro sagittal saw attachment, was being used during a laterjet procedure on (b)(6) 2023 and "the powerpro sagittal saw attachment was being used with a 90 degree saw blade.When applying force the cut, the blade came loose and shot out of the hub nearly hitting a staff member¿.The procedure was completed using a pro8250sb, and there was a delay of ten minutes.There was no report of injury to the patient or user.A good faith effort was made to obtain additional information; no response was received.Update: per information received on 29nov23, the blade was fully locked in place at the time of the event, and there was no fragmentation.This report is being raised due to the reported malfunction with potential for injury upon reoccurrence.
|
| |
|
Additional Manufacturer Narrative
|
|
The device will not be returned and no photographic evidence has been provided therefore the reported event cannot be verified.The service history was reviewed, and no data was found.A device history record review found no abnormalities that would contribute to the reported event.A two-year review of complaint history revealed there has been a total of 11 reports, regarding 11 devices, for this device family and failure mode.During this same time frame (b)(4) devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be (b)(4).Per the instructions for use, the user is advised: prior to each use, inspect all equipment for proper operation and ensure all attachments and accessories are correctly and completely attached to the handpiece.Handle all equipment carefully.If an attachment is dropped or damaged in any way, return it immediately for service.We will continue to monitor for trends through the complaint system to assure patient safety.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: information has been updated in sections b.5 and d.10, identifying a concomitant device.Correction: brand name has been corrected in section d.1.Manufacturer narrative: the device will not be returned and no photographic evidence has been provided therefore the reported event cannot be verified.The service history was reviewed, and no data was found.A device history record review found no abnormalities that would contribute to the reported event.A two-year review of complaint history revealed there has been a total of 11 reports, regarding 11 devices, for this device family and failure mode.(b)(4).Per the instructions for use, the user is advised: prior to each use, inspect all equipment for proper operation and ensure all attachments and accessories are correctly and completely attached to the handpiece.Handle all equipment carefully.If an attachment is dropped or damaged in any way, return it immediately for service.We will continue to monitor for trends through the complaint system to assure patient safety.
|
| |
|
Event or Problem Description
|
|
A sales representative reported on behalf of a customer that the device, pro2043, powerpro sagittal saw attachment, was being used during a laterjet procedure on (b)(6) 2023 and "the powerpro sagittal saw attachment was being used with a 90 degree saw blade.When applying force the cut, the blade came loose and shot out of the hub nearly hitting a staff member¿.The procedure was completed using a pro8250sb, and there was a delay of ten minutes.There was no report of injury to the patient or user.A good faith effort was made to obtain additional information; no response was received.Update: per information received on 29nov23, the blade was fully locked in place at the time of the event, and there was no fragmentation.Update #2: per information received on 25jul24, the 90 degree saw blade has been identified as the 00502313600, micro sagittal blade, coarse.There are no allegations made against it.This report is being raised due to the reported malfunction with potential for injury upon reoccurrence.
|
| |
|