ZIMMER BIOMET, INC. 36MM ÿ +6MM OFFSET POLY LINER; PROSTHESIS, SHOULDER Back to Search Results
Catalog Number 00434903606
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/09/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a poly liner would mate not with a tm stem spacer upon impaction.Another spacer and liner were used to complete the procedure.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
(b)(4).D10: 9mm humeral stem spacer cat# 00434903909, lot# 64147467 product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2023-03314.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual examination of the returned product/provided pictures, identified the anti-rotation slot on the poly liner was measured, using a plug gauge method.The feature is conforming to print specification.Damage or witness mark is seen on the backside of the slot.The device was conforming to print specifications.Review of the device history record(s) identified no deviations or anomalies, during manufacturing related to the reported event.A definitive root cause cannot be determined.The reported event is confirmed, by returned product.If any further information is found, which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
36MM ÿ +6MM OFFSET POLY LINER
Common Device Name
PROSTHESIS, SHOULDER
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18227589
Report Number0001822565-2023-03313
Device Sequence Number20384140
Product Code PHX
UDI-Device Identifier00889024269071
UDI-Public(01)00889024269071(17)280919(10)64687458
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K052906
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number00434903606
Device Lot Number64687458
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/15/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/09/2023
Supplement Date Received by Manufacturer05/16/2024
Initial Report FDA Received Date11/29/2023
Supplement Report FDA Received Date05/17/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/25/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-