• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. 7" (18 CM) APPX 0.52 ML, SMALLBORE TRIFUSE EXT SET W/3 MICROCLAVE® CLEAR, 3 CLAM; STOPCOCK, I.V. SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. 7" (18 CM) APPX 0.52 ML, SMALLBORE TRIFUSE EXT SET W/3 MICROCLAVE® CLEAR, 3 CLAM; STOPCOCK, I.V. SET Back to Search Results
Catalog Number MC330027
Medical Device Problem Codes Crack (1135); Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/06/2023
Type of Reportable Event Malfunction
Event or Problem Description
The incident involved a 7" (18 cm) appx 0.52 ml, smallbore trifuse ext set w/3 microclave¿ clear, 3 clamps, luer lock.It was stated in the report that the trifurcate had a micro-clave that was cracked, it caused blood to flow back to the peripherally inserted central catheter (picc) line and for the fats to leak out.When it was time for the healthcare professional to do the assessment, it was noticed that the bed was soaked.Upon tracing back the lines, it was discovered that the microclave had a crack in it.There was patient involvement and no patient harm reported.
 
Additional Manufacturer Narrative
The device has been requested to be returned for evaluation; however, it has not yet been received.
 
Additional Manufacturer Narrative
No product samples, videos, or photographs were provided for investigation.The device history review (dhr) was reviewed and there were no relevant non-conformances found that would have contributed to the reported complaint.A probable cause cannot be identified based on the information that has been provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
7" (18 CM) APPX 0.52 ML, SMALLBORE TRIFUSE EXT SET W/3 MICROCLAVE® CLEAR, 3 CLAM
Common Device Name
STOPCOCK, I.V. SET
Manufacturer (Section D)
ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V.
avenida cuarzo no. 250
ensenada, b.cfa. 22790
MX  22790
Manufacturer Contact
reed covert
600 n. field dr.
lake forest, IL 60045
2247062300
MDR Report Key18228825
Report Number9617594-2023-01043
Device Sequence Number6212435
Product Code FMG
UDI-Device Identifier00887709076419
UDI-Public(01)00887709076419(17)280601(10)13661842
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K964435
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberMC330027
Device Lot Number13661842
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/17/2023
Supplement Date Received by Manufacturer12/19/2023
Initial Report FDA Received Date11/29/2023
Supplement Report FDA Received Date12/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/01/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PICC LINE, MFR UNK.; UNSPECIFIED FAT INFUSION, MFR UNK.
Patient SexUnknown
-
-