| Catalog Number |
TSVM4B13 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); Osteolysis (2377)
|
| Date of Event |
09/18/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Zimvie complaint (b)(4).A2: patient age is not provided / unknown.A4: patient weight is not provided / unknown.E1: initial reporter¿s title is not provided / unknown.
|
| |
|
Event or Problem Description
|
|
It was reported that the implant was removed due to peri-implantitis.Tooth site # 15.Site grafted at time of implant placement with allograft and cortical genet blend.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to report corrected information to 0002023141-2023-03418.Customer confirmed that there was no peri implantitis present.
|
| |
|
Event or Problem Description
|
|
It was reported that the implant was removed due to infection and bone loss.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to report additional information.A summary investigation has been completed for bone loss and infection events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for bone loss and infection have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of manufacturing or design defects that might lead to bone loss and escaping the available detections has been assessed and found remote and almost nonexistent.Dhr review was completed for the subject lot number.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Sterilization record was reviewed and verified to have passed all sterilization activities with no issues or nonconformities identified.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.As documented in the summary investigations, contributing factors for the reported event likely exist outside of zimvie control, including those related to patient biological factors/condition and surgical technique.Based on the summary investigations, no malfunction occurred upon investigation.The reported events remain non-verifiable as they are a medical condition.Monthly trending to date has not identified any statistical triggers suggesting potential design or manufacturing related issues.Previously completed investigations for infection and/or bone loss, problems reported in association with implant failure have not identified any signals indicating potential non-conformances impacting the manufacturing and sterilization processes.Therefore, escalation to capa is not required.
|
| |
|
Event or Problem Description
|
|
No additional or corrected information to report.
|
| |
|
Search Alerts/Recalls
|