Model Number 233-150
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/26/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the anterolateral inserter was not working properly.
 
Additional Manufacturer Narrative
Visual inspection of the device confirmed the distal arm had detached and separated from the shaft.The distal arm was not returned with the device.The distal arm of the inserter is a subcomponent which features an ipsilateral prong that attaches to the interbody spacer during insertion.This type of damage may be caused when force is applied during impaction/insertion of the interbody spacer causing the distal arm to detach from the shaft.Review of the device history record indicates the device conformed to dimensional specifications at the time of manufacturing with no non-conformities reported.Labeling review: "warnings/cautions/precautions: potential risks identified with the use of these fusion devices, which may require additional surgery, include device component failure, loss of fixation, pseudoarthrosis (i.E.Non-union), fracture of the vertebra, neurological injury, and/or vascular or visceral injury." "possible adverse effects: possible adverse effects include: 1.Initial or delayed loosening, bending, dislocation, and/or breakage of device components.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IDENTITI ALIF
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH INTEGRATED FIXATION, LUMBAR
Manufacturer (Section D)
ALPHATEC SPINE, INC.
1950 camino vida roble
carlsbad CA 92008
Manufacturer (Section G)
ALPHATEC SPINE, INC.
1950 camino vida roble
carlsbad CA 92008
Manufacturer Contact
wesley channell
1950 camino vida roble
carlsbad, CA 92008
9014283693
MDR Report Key18232899
Report Number2027467-2023-00079
Device Sequence Number15199532
Product Code OVD
UDI-Device Identifier00190376382982
UDI-Public(01)00190376382982(10)EM46681
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K203742
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 11/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number233-150
Device Catalogue Number233-150
Device Lot NumberEM46681
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/30/2023
Initial Report FDA Received Date11/29/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/03/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-