| Catalog Number |
383336 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/23/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H.3.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that bd saf-t-intima w/y adapter pnk 20gax1.0in needle broke after successful puncture of the indwelling needle, the guidewire was broken when retiring the guidewire during tube placement, and part of the guidewire was retained in the indwelling needle; the retained broken guidewire was later removed with forceps and reported without adverse consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
A device history record review was completed by our quality engineer team for provided material number 383336 and lot number 2066440.The review did not reveal any detected abnormalities during the production process that could have contributed to this defect and all quality tests were found to be within specification.As a sample was unavailable for return, a thorough sample investigation could not be completed.Based on the investigation results, an exact cause for this incident could not be identified.Should you again experience any problems with our product we would appreciate the opportunity to conduct a thorough analysis.There are quality controls currently in place to detect this type of defect during the production process.Further action has not been determined necessary at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Our quality team regularly reviews the collected data for identification of emerging trends.H3 other text : see narrative below.
|
| |
|
Search Alerts/Recalls
|