| Catalog Number |
4252543-02 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
As reported by the user facility: reported issue per customer: "recently there have been complaints that even when tightening the iv catheter to the iv tubing there is leaking of the fluids, blood, and even the meds for sedation." no injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.Corrected data: please note that this report is being resubmitted with the original submission date of 13oct2023 as there was an issue during the initial submission.
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).A review of the batch history records was performed for the reported lot number and no non-conformances or deviations were noted during the manufacturing process or final inspections.No sample was provided for evaluation.Based on the data from the investigation, we are unable to determine the root cause of the reported incident.The reported defect was unable to be confirmed.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|
| |
|