• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ERIKA DE REYNOSA, S.A. DE C.V. LIBERTY CYCLER SET, SINGLE PATIENT CONN; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ERIKA DE REYNOSA, S.A. DE C.V. LIBERTY CYCLER SET, SINGLE PATIENT CONN; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 050-87215
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Abdominal Pain (1685); Peritonitis (2252)
Date of Event 11/08/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Clinical review: a temporal relationship exists between the pd therapy with the liberty select cycler/ liberty cycler set and the patient¿s peritonitis event.Based on the reported information, the cause of the patient¿s peritonitis cannot be determined at this time.However, currently there is no allegation nor any objective evidence that a liberty cycler set malfunction or product deficiency was associated with this event.Peritonitis is a well-documented complication in patients undergoing pd therapy most often associated with a breach in aseptic technique during the pd exchange.The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
 
Event or Problem Description
A patient's contact reported that this patient on pd therapy has peritonitis.There were no reported allegations that the peritonitis was related to any issues with products.In additional follow-up, the pd nurse stated the patient was having symptoms of abdominal pain and cloudy pd effluent.As a result, on (b)(6) 2023, the patient was seen in the outpatient pd clinic.The patient had a pd culture obtained which grew candida tropicalis.The nurse stated the patient is being treated with antibiotic therapy using oral fluconazole (dose unknown) and intravenous vancomycin and ceftazidime (dose unknown) on an outpatient basis.Additionally, the patient was transitioned to back up hemodialysis for renal replacement needs and will undergo removal of the pd catheter (not a fresenius product).The patient¿s nurse was not able to identify the exact cause of the patient¿s peritonitis is unknown.However, the nurse confirmed the patient did not have any fluid leaks or issues with fresenius products in relation to this event.The nurse reported the patient is recovering from peritonitis event.
 
Additional Manufacturer Narrative
Plant investigation: the sample was not returned to the manufacturer and the lot number was not provided.A manufacturing review was performed on the products shipped to the patient for the three (3) month time frame which immediately preceded the event occurrence date.This review included the lot numbers for all fresenius liberty cycler sets shipped to this account within the selected time frame.The entire set of lots have been sold and distributed.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.An investigation of the device history records (dhr) was conducted and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The product lots involved met all specifications for release.A review of the dhr did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
 
Event or Problem Description
A patient's contact reported that this patient on pd therapy has peritonitis.There were no reported allegations that the peritonitis was related to any issues with products.In additional follow-up, the pd nurse stated the patient was having symptoms of abdominal pain and cloudy pd effluent.As a result, on (b)(6) 2023, the patient was seen in the outpatient pd clinic.The patient had a pd culture obtained which grew candida tropicalis.The nurse stated the patient is being treated with antibiotic therapy using oral fluconazole (dose unknown) and intravenous vancomycin and ceftazidime (dose unknown) on an outpatient basis.Additionally, the patient was transitioned to back up hemodialysis for renal replacement needs and will undergo removal of the pd catheter (not a fresenius product).The patient¿s nurse was not able to identify the exact cause of the patient¿s peritonitis is unknown.However, the nurse confirmed the patient did not have any fluid leaks or issues with fresenius products in relation to this event.The nurse reported the patient is recovering from peritonitis event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIBERTY CYCLER SET, SINGLE PATIENT CONN
Common Device Name
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
director, quality systems
900 w zaragosa drive suite d
pharr TX 78577
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
6174175172
MDR Report Key18234119
Report Number0008030665-2023-01095
Device Sequence Number11569587
Product Code FKX
UDI-Device Identifier00840861100743
UDI-Public00840861100743
Combination Product (Y/N)N
Initial Reporter StateVT
Initial Reporter CountryUS
PMA/510(K) Number
K173718
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue Number050-87215
Was Device Available for Evaluation? No
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/13/2023
Supplement Date Received by Manufacturer01/04/2024
Initial Report FDA Received Date11/29/2023
Supplement Report FDA Received Date01/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
LIBERTY CYCLER SET.; LIBERTY CYCLER SET.; LIBERTY SELECT CYCLER.; LIBERTY SELECT CYCLER.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age79 YR
Patient SexMale
-
-