| Model Number |
TRULIANT TIB IMP PS INSERT SZ 3.5 9MM |
| Medical Device Problem Code |
Device Markings/Labelling Problem (2911)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/06/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
As reported, during surgical use, when the staff opened the 3.5-9mm ps insert, they found that a 5-9mm was inside.A ps insert 3.5-10mm was inserted.There was no breakage of device.There was a greater than five minutes surgical delay; however, the patient did not experience any adverse events as a result of delay.Images received.The device will be return.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The complaint device was evaluated, and the reported event was confirmed.The evaluation identified the packaging as labeled did not match the device as manufactured.A definitive root cause was unable to be determined at this time; however, further investigation into the reported event is being conducted through our capa process.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Search Alerts/Recalls
|