• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH DELTA V-40 CERAMIC HEAD 36/0; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH DELTA V-40 CERAMIC HEAD 36/0; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS Back to Search Results
Catalog Number 6570-0-136
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Unequal Limb Length (4534)
Date of Event 06/01/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Incident report received by the tga 15/09/2023.I'm concerned as this product was alerted by stryker in jan 2022 as a voluntary recall.I'm wondering if australia was issued the same alert? there are numerous documents that i have found online by stryker stating the product of this lot number was possibly defective and wrong sizing compared to what stated on product packaging.There were numerous complications mentioned in the alerts that it cause leg length discrepancy, weakness of the abductor muscle, hip dislocation of femoral head from acetabular cup, and pain and inability to walk leading to revision surgery.My partner had this device put in him last year on (b)(6) 2022 and had all these issues plus more.The pain and suffering has been over 15 months and very poor management post care from the surgeon which is really poor duty of care.
 
Additional Manufacturer Narrative
Reported event: an event regarding limb length discrepancy involving a ceramic head was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.The patient would like to know if the implant was subject to a recall.Upon review, there are no recalls associated with this device.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.H3 other text: device not returned to the manufacturer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DELTA V-40 CERAMIC HEAD 36/0
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
arokiya raj
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key18234583
Report Number0002249697-2023-01470
Device Sequence Number19640266
Product Code LZO
UDI-Device Identifier04546540608512
UDI-Public04546540608512
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K052718
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number6570-0-136
Device Lot Number93950103
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/06/2023
Initial Report FDA Received Date11/29/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/05/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
-
-