| Catalog Number |
011-C1000 |
| Medical Device Problem Codes |
Disconnection (1171); Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/02/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is not available for investigation.However, photographs was provided by the customer and evaluation of the photos are pending.
|
| |
|
Event or Problem Description
|
|
The complaint/event involved a clave® connector.The head of oncology at the hospital reported that the set disassembled during treatment (after approximately one hour).Microdroplets in pink color (doxorubicin) was observed on the membrane side.The problem was noticed during handling at the moment of disconnection after treatment.Although there was a patient involved, there was no delay in therapy, no adverse event and no one harmed as a result of this event.Neither the patient nor the clinical professional were in contact with the fluid, because the incident was detected and the device was disposed of using protective measures.
|
| |
|
Additional Manufacturer Narrative
|
|
Two photos were shared by the customer.The first photo illustrated a microclave with small drops coming from the seal.However, the pink color described by customer was not observed.The second photo showed the label information with lot number and item description.No additional damages were observed.The device history record was reviewed and no non conformities were found that would have led the reported condition on the complaint.Additional information can be found in h6 medical device problem code.
|
| |
|