MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS Back to Search Results
Model Number ZCB00
Medical Device Problem Codes Device Difficult to Setup or Prepare (1487); Operating System Becomes Nonfunctional (2996)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/31/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information: sections a-2 patient age/date of birth, a-4 patient weight, and a-5 patient ethnicity and race: unknown/asked information unavailable.Section d-6a date implanted: not applicable as the iol was removed and replaced during the initial procedure.Section d-6b date explanted: not applicable as the iol was removed and replaced during the initial procedure, therefore not explanted.Section h3-81: the device was not returned for evaluation.Therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported upon implantation the zcb00 model intraocular lens (iol) was damaged, there was a lot of resistance while trying to inject and the plunger wouldn't advance correctly.The iol was cut out and the procedure was completed using a back-up lens with the same diopter.There was no patient injury and the case proceeded as planned.Patient status post-procedure was reported as doing well.No further information is available.
 
Additional Manufacturer Narrative
Additional information: section d-9: device available for evaluation: yes.Section d-9: device date returned to manufacturer: 08-dec-2023.Section h-3: device evaluated by manufacturer: yes.Device evaluation: the complaint lens was received in a bag with the complaint lens case.The complaint lens was visually inspected under magnification.The complaint lens was received loose in a bag.The lens was cleaned and no issues that could have contributed to the complaint event were identified.Complaint issue "difficult to use" and "plunger rod issue" were not identified during product evaluation.Therefore, there is no indication of a product deficiency or product malfunction.Conclusion: based on the complaint investigation results, the product was released within specifications and the reported complaint issues could not be confirmed.No product deficiency or product malfunction was identified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key18235111
Report Number3012236936-2023-03031
Device Sequence Number14355847
Product Code HQL
UDI-Device Identifier05050474531741
UDI-Public(01)05050474531741(17)260408
Combination Product (Y/N)N
Initial Reporter StateMT
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/08/2026
Device Model NumberZCB00
Device Catalogue NumberZCB0000215
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/07/2023
Supplement Date Received by Manufacturer12/15/2023
Initial Report FDA Received Date11/29/2023
Supplement Report FDA Received Date12/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/08/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-