| Catalog Number |
ASK-44552-ABKB |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/06/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that: the handle of the peel away sheath broke off.The reported defect was detected during use on patient.The patient condition was reported as "fine".No patient injury/consequence or medical intervention reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).The customer report of a broken peel away sheath tab was confirmed by visual inspection of the returned sample and customer photos.The customer provided two photos and returned one peel away sheath for analysis.Definite signs of use were observed.Visual analysis revealed that one peel away sheath extrusion was separated directly adjacent to the tab.Remains of the extrusion was still within the separated tab.The other tab was intact.It was also noted that the sheath was entirely peeled down both score lines as intended.The overall length of the sheath measured 2 3/4" via calibrated ruler which was within the specification limits of 2 5/8" - 2 7/8" per the sheath product drawing.The outer and inner diameter of the sheath could not be performed due to the condition of the returned sample.Functional inspection could not be performed due to the condition of the returned sample.The instructions for use (ifu) provided with this kit warns the user, "do not use excessive force when introducing guidewire, peel-away sheath over tissue dilator, or tissue dilator as this can lead to vessel perforation and bleeding, or component damage." a device history record review was performed, and a relevant finding was identified.Based on the customer report, sample received, and comments from manufacturing, the potential root cause of this event is manufacturing related.Further investigation has been initiated under teleflex's quality system by the manufacturing site to further investigate this issue.Teleflex will continue to monitor and trend on reports of this nature.
|
| |
|
Event or Problem Description
|
|
It was reported that: the handle of the peel away sheath broke off.The reported defect was detected during use on patient.The patient condition was reported as "fine".No patient injury/consequence or medical intervention reported.Associated to 9680794-2023-00950.
|
| |
|
Search Alerts/Recalls
|