| Model Number |
5430-30152 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Thyroid uptake system s/n (b)(6) was shipped on 2/23/2016.The technician was performing a thyroid uptake on patient when the arm unexpectedly fell to its lowest position.Review by engineering indicated a spring arm failure.No injuries occurred.A replacement assembly was shipped and user instructed to return the defective part to manufacturer for engineering evaluation and identification of root cause.The collimator and arm weigh and spring arm assembly weigh 45 pounds, which has the potential for serious injury to either patient or operator.
|
| |
|
Additional Manufacturer Narrative
|
|
Safety alert sent to installed base of potentially suspect units.Additional corrective actions pending completed engineering analysis.
|
| |
|