MORTARA INSTRUMENT ELI 380; ELECTROCARDIOGRAPH Back to Search Results
Model Number ELI380-DCS11
Medical Device Problem Code Wireless Communication Problem (3283)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/02/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The eli 380 device is indicated for use to acquire, analyze, display, and print electrocardiograms.Device is indicated for use to provide interpretation of the data for consideration by a physician.Device is indicated for use in a clinical setting, by a physician or by trained personnel who are acting on the orders of a licensed physician.It is not intended as a sole means of diagnosis.The interpretations of ecg offered by the device are only significant when used in conjunction with a physician over-read as well as consideration of all other relevant patient data.Device is indicated for use on adult and pediatric populations.The device is not intended to be used as a vital signs physiological monitor.The device was returned for repair where the reported issues were verified.The device failed the functional check.The problem was determined to be with the motherboard and keyboard.The hrc technician replaced the motherboard, keyboard/upper housing, utk and replaced the monitors to resolve the reported problems.The software was also updated to version 2.6.1.1.And the customer was advised to reset their wlan and lan settings.Based on this information, no further actions are required at this time.If the eli 380 device loses connection to the wireless network, the device is unable to transmit or receive ecg¿s.This failed connection may result in a delay of patient treatment which may cause or contribute to a serious injury or death.Therefore, hillrom is reporting this connection failure as a product malfunction.
 
Event or Problem Description
The customer reported a number of issues with the eli380 including the wired and wireless network connections would simply stop working but then come back with no action taken.The customer also stated that when any of those conditions occur, the screen goes blank.There was no associated adverse event reported with this complaint.This incident was captured under hill-rom complaint ref # (b)(4).
 
Additional Manufacturer Narrative
This supplemental report is a correction to the previously submitted mdr.Following communication from the fda regarding incorrect/missing udi numbers, baxter reviewed the subject mdr and determined brand name, 510k, and udi corrections were required to this mdr in alignment with the gudid.
 
Event or Problem Description
The customer reported a number of issues with the eli380 including the wired and wireless network connections would simply stop working but then come back with no action taken.The customer also stated that when any of those conditions occur, the screen goes blank.There was no associated adverse event reported with this complaint.This incident was captured under baxter complaint ref (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELI 380
Common Device Name
ELECTROCARDIOGRAPH
Manufacturer (Section D)
MORTARA INSTRUMENT
7865 n 86th street
milwaukee WI 53224
Manufacturer Contact
kayla miller
1069 state route 46 east
batesville, IN 47006
8129310130
MDR Report Key18236782
Report Number2183461-2023-00026
Device Sequence Number7160058
Product Code DPS
UDI-Device Identifier00812345026917
UDI-Public(01)00812345026917
Combination Product (Y/N)N
Initial Reporter StateVT
Initial Reporter CountryUS
PMA/510(K) Number
K142105
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberELI380-DCS11
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/02/2023
Supplement Date Received by Manufacturer11/02/2023
Initial Report FDA Received Date11/30/2023
Supplement Report FDA Received Date06/17/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-