• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. Z.S.G.M. CUTTER 4:1 RATIO ( CAUTION: SHARP ); EXPANDER, SURGICAL, SKIN GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. Z.S.G.M. CUTTER 4:1 RATIO ( CAUTION: SHARP ); EXPANDER, SURGICAL, SKIN GRAFT Back to Search Results
Model Number N/A
Medical Device Problem Codes Material Twisted/Bent (2981); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/11/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This event is recorded with zimmer biomet under (b)(4) this medwatch is being filed as an initial / final report based on information discovered during the device evaluation.Review of the most recent repair record identified the following related repair: the cutter failed testing due to an incomplete cut.The cutter will need replacement.Review of the device history records identified no deviations or anomalies during manufacturing.The root cause of the reported issue is attributed to design issue.The event is confirmed.Additional related reports: 0001526350-2023-00715-1.0001526350-2023-01562.0001526350-2023-01564.0001526350-2023-01565.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that the 2:1 roller was tight.When the dr.Checked the instrument, they noticed with grill was bent.They tried to unbend it.They managed to use it for the case but noticed the ratcheting handle was also not working like it was supposed to.Event occurred prior to surgery.There was no reported patient harm, or delay.Due diligence is complete, no further information is available at this time.The evaluation found the cutter failed the test cut.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Z.S.G.M. CUTTER 4:1 RATIO ( CAUTION: SHARP )
Common Device Name
EXPANDER, SURGICAL, SKIN GRAFT
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key18238535
Report Number0001526350-2023-01566
Device Sequence Number6936192
Product Code FZW
UDI-Device Identifier00889024375475
UDI-Public(01)00889024375475(11)181016(10)64187517
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number00770304000
Device Lot Number64187517
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/22/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/27/2023
Initial Report FDA Received Date11/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/23/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
00770102200.RATCHET HANDLE.
Patient SexPrefer Not To Disclose
-
-