| Model Number |
8881512977 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/07/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that when provider uses syringe to administer medication, it is coming out of the top side of syringe (no proper seal on syringe).Additional information received from the initial reporter on 15nov2023 stated that it appears the plunger does not seal to the syringe and fluid is able to bypass.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the information available to us, we were unable to confirm the event.The device history record review found no abnormal process conditions present during manufacturing that could have led to the reported issue.Samples were received and visually evaluated; the samples met product specification, there were no visually issues found.The samples were forwarded to the supplier for further review.In the meantime, all information received will be used for further tracking and trending purposes.
|
| |
|