| Model Number |
N160 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Failure to Capture (1081); Over-Sensing (1438); Pacing Problem (1439); Under-Sensing (1661)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/24/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during the explant procedure of this cardiac resynchronization therapy defibrillator (crt-d), the patient exhibited oversensing from electromagnetic interference (emi) from the electrocautery, leading into pacing inhibition, undersensing and loss of capture (loc).Technical services (ts) explained to the field representative that this is indicative of defibrillator detect circuit behavior.No asystole was exhibited.This crt-d remains in service.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Visual examination of the device header and case noted no anomalies.Dimensional analysis of the header was completed.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the pacing, and sensing functions were tested.Impedance testing was completed, and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations outside the bounds of normal medical use.
|
| |
|
Event or Problem Description
|
|
It was reported that, during the explant procedure of this cardiac resynchronization therapy defibrillator (crt-d), the patient exhibited oversensing from electromagnetic interference (emi) from the electrocautery, leading into pacing inhibition, undersensing and loss of capture (loc).Technical services (ts) explained to the field representative that this is indicative of defibrillator detect circuit behavior.No asystole was exhibited.No adverse patient effects were reported.
|
| |
|