| Model Number |
BT-80-0004 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/03/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Upon troubleshooting it was determined the power cord needed to be replaced.Physical damage of a cord would be readily noticeable to the user.The user would notice any integrity issues with the surrounding insultation (outer covering) when plugging or unplugging the device into the electrical outlet.If found to be damaged, the power cord and device would be immediately removed from use.The power supply cords, by design, meet (iec 60601-1-6 usability standard), (iec 60245-1:2003, annex a, designation 53) or (iec 60227-1:1993, annex a, designation 53).Additionally, specific electrical safety standards apply to each product and subsequent applicable usability testing.A search of the baxter maintenance records did not show baxter performed any preventative maintenance on this device.A power cord replacement was shipped direct to patient to resolve the reported event.Based on this information, no further action is required.
|
| |
|
Event or Problem Description
|
|
It was reported the power cord was damaged with exposed copper wires.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported the power cord was damaged with exposed copper wires.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is a correction to the previously submitted mdr.Following communication from the fda regarding incorrect/missing udi numbers, baxter reviewed the subject mdr and determined brand name and udi corrections were required to this mdr in alignment with the gudid.
|
| |
|