MEDOS INTERNATIONAL SARL TRUFILL N-BCA-1 GRAM KIT; TISSUE ADHESIVE FOR USE IN EMBOLIZATION OF BRAIN ARTERIOVENOUS MALFORMATIONS Back to Search Results
Catalog Number 631500
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Stroke/CVA (1770); Confusion/ Disorientation (2553); Convulsion/Seizure (4406)
Date of Event 11/10/2023
Type of Reportable Event Serious Injury
Event or Problem Description
As reported via the membrane study, an 82-year-old female (subject (b)(6)) with a medical history of coronary artery disease and hypertension (controlled), underwent middle meningeal artery (mma) embolization of a left frontal subdural hematoma (sdh) with temporal and parietal involvement on (b)(6) 2023.The patient was randomized into the non-surgical and mma embolization cohort of the study.The patient¿s baseline markwalder grading scale (mgs) score was 0 and the modified rankin scale (mrs) score was 0.The sdh thickness was 11.The mma embolization study procedure was performed on (b)(6) 2023 via right femoral access.Trufill n-bca-1 gram kit (631500/m4378t) (4:1 oil: n-bca ratio) was delivered to midline via a headway duo (microvention).The anterior and posterior branches of the mma were embolized.In the opinion of the treating physician, the embolization was successful.The event of ¿increased confusion,¿ which started on 10-nov-2023, was assessed by the pi as a serious, severe adverse event, and as not related to the study device, possibly related to the mma embolization procedure, not related to the medication used to treat the subdural hematoma (sdh), and possibly related to the surgical procedure used to treat the sdh.The event required the diagnostic test of a ct brain.The outcome is recorded as ¿not recovered/not resolved.¿ the event of ¿punctate acute stroke left corona radiata,¿ which started on (b)(6) 2023, was assessed by the pi as a serious, severe adverse event, and as not related to the study device, possibly related to the mma embolization procedure, not related to the medication used to treat the subdural hematoma (sdh), and possibly related to the surgical procedure used to treat the sdh.The event was treated with medication.The outcome is recorded as ¿recovering/resolving.¿ the patient was discharged from the hospital on (b)(6) 2023 with a mini-mental state exam (mmse) score of 8, a markwalder grading scale (mgs) score of 2, and a modified rankin scale (mrs) score of 4.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Device manufacture date: not available at the time of report.Stroke and neurological deficits, such as confusion, are both known potential complications associated with the use of the trufill n-bca liquid embolic agent and is mentioned in the instructions for use (ifu) as such.There were no alleged quality issues/ malfunctions related to the device, as the device performed as intended.The events of ¿increased confusion and punctate acute stroke left corona radiata¿ were both assessed by the pi as unrelated to the study device but possibly related to the mma embolization study procedure and possibly related to the surgical procedure used to treat the sdh.In patients of advanced age (82-year-old), increased confusion can be attributed to the urinary tract infection (uti), however, should be resolved once the infection resolves.In this case, it was reported that the uti resolved with an end date of (b)(6) 2023, while the confusion did not.The events were assessed as possibly related to the mma embolization study procedure (with n-bca), and the relationship of the device to the reported events, although remote, based on the principal investigator's assessment of not related, cannot be excluded as a factor in the events.Therefore, the events of stroke and confusion will be conservatively reported to the us fda reporting under criteria 21 cfr 803 with a classification of ¿serious injury.¿ as part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.Reference: dutta c, pasha k, paul s, abbas ms, nassar st, tasha t, desai a, bajgain a, ali a, mohammed l.Urinary tract infection induced delirium in elderly patients: a systematic review.Cureus.2022 dec 8;14(12):e32321.Doi: 10.7759/cureus.32321.Pmid: 36632270; pmcid: pmc9827929.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).The device remains implanted; therefore, no further investigation can be performed.A manufacturing record evaluation was performed for the finished device m4378t number, and no non-conformances related to the malfunction were identified.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Section b5: additional/modified information was received on 10-jan-2024.Summary: regarding the adverse event of ¿increased confusion,¿ the severity value of "severe" has been changed to "moderate." a supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.
 
Additional Manufacturer Narrative
Manufacturer ref#: (b)(4).Updated sections on this medwatch: a1, a2, b4, b5, d4 (primary udi number) g3, g6, h2, h6 ( health effect - clinical code) and (health effect - impact code) and h11.On 23-aug-2024, a clinical event committee (cec) adjudication for the adverse event of ¿punctate acute stroke left corona radiata¿ was received.The relationship of event to the mma embolization procedure and surgical procedure used to treat the sdh were assessed as ¿possibly related.¿ the relationship of event to the study device and sdh medication, were assessed as ¿not related.¿ on 28-aug-2024, a clinical event committee (cec) adjudication for the adverse event of ¿bitemporal seizure, multiple seizures over time¿ was received.The relationship of event to the study device and mma embolization procedure were assessed as ¿possibly related.¿ the relationship of event to the surgical procedure used to treat the sdh and sdh medication, remains as ¿not related.¿ it was further commented, ¿because glue was used in this procedure it could have caused irritation to cause a seizure.¿ additionally, a clinical event committee (cec) adjudication for the adverse events of ¿worsening symptomatic subdural hematoma and right posterior cerebral artery aneurysm rupture¿ were received.The events remain assessed as not related to the study device, the mma embolization procedure, the medication used to treat the subdural hematoma (sdh), nor to the surgical procedure used to treat the sdh.Per the additional/modified information received on 28-aug-2024; seizures are a known potential complication associated with the use of the trufill n-bca liquid embolic agent and is mentioned in the instructions for use (ifu) as such.There was no alleged quality issues related to the device, as the device performed as intended.The principal investigator assessed this event as unrelated to the study device; however, the clinical event committee (cec) adjudication deemed the event ¿possibly related¿ to the study device and to the mma study procedure.Based on this assessment, the correlating relationship between the event to the used trufill n-bca study device cannot be ruled out entirely.Additionally, the event was deemed a serious adverse event, which required in-patient hospitalization, diagnostic evaluations and medicinal treatment.Based on the assessment of the clinical event committee (cec), this event does meet us fda reporting criteria under 21 cfr 803 with a classification of ¿serious injury,¿ with an awareness date of 28-aug-2024.The file will be re-reviewed if additional information is received at a later date.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUFILL N-BCA-1 GRAM KIT
Common Device Name
TISSUE ADHESIVE FOR USE IN EMBOLIZATION OF BRAIN ARTERIOVENOUS MALFORMATIONS
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel CH-24 00
SZ  CH-2400
Manufacturer (Section G)
CERENOVUS INC
325 paramount dr
raynham MA 02767
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key18239745
Report Number3008114965-2023-00878
Device Sequence Number3240529
Product Code KGG
UDI-Device Identifier10886704029151
UDI-Public(01)10886704029151(17)250131(10)M4378T
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
PMA/510(K) Number
P990040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 04/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/31/2025
Device Catalogue Number631500
Device Lot NumberM4378T
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/06/2023
Supplement Date Received by Manufacturer12/11/2023
01/10/2024
08/28/2024
Initial Report FDA Received Date11/30/2023
Supplement Report FDA Received Date12/12/2023
01/11/2024
04/25/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/25/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization; Life Threatening;
Patient Age82 YR
Patient SexFemale
Patient EthnicityNon Hispanic