• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC COMPACT SPEED REDUCER, 60:1; MOTOR, DRILL, PNEUMATIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC COMPACT SPEED REDUCER, 60:1; MOTOR, DRILL, PNEUMATIC Back to Search Results
Catalog Number CSR60
Medical Device Problem Code Excessive Heating (4030)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/19/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: d10: concomitant med products and therapy dates: motor device, 11/19/2023.E1: the reporter¿s phone number was not provided.Device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed, and it was determined that the reported condition of the device generating heat was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation, it was determined that the device had component damage, worn bearing, vibration, unexpected noise and was difficult to assemble/disassemble.It was further determined that the device failed pretest for visual assessment.The assignable root cause was determined to be due to component failure from normal wear.Udi: (b)(4).
 
Event or Problem Description
This is report 1 of 2 of the same event: it was reported from japan that the motor device and compact speed reducer device were generating heat while being used together.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMPACT SPEED REDUCER, 60:1
Common Device Name
MOTOR, DRILL, PNEUMATIC
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
kate karberg
4500 riverside drive
palm beach gardens, FL 33410
3035526892
MDR Report Key18239954
Report Number1045834-2023-01657
Device Sequence Number3225194
Product Code HBB
UDI-Device Identifier00845384001782
UDI-Public00845384001782
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K960630
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberCSR60
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/20/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/19/2023
Initial Report FDA Received Date11/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/27/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-