| Catalog Number |
L-70NI |
| Medical Device Problem Codes |
Break (1069); Material Separation (1562); Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that due to damage to the lure portion of the connection, it could not be connected to the route and its use was discontinued.No patient injury or clinical affects was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Other text: b3: month and year of event have been provided, day is unknown.D4: lot number, expiration date, and h4: device manufacture date is unknown, no information has been provided to date.D3, g1,g2 email is:(b)(6) investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: device evaluation: one sample was received in used condition.A visual inspection found a crack in the female luer connector.During the functional testing, a leak test was done and during the third attempt the female luer presented damage in the connection part.However, the damage was not similar to the sample reported.Based on the replication of the failure mode results, the crack reported is not attributable to the manufacturing process.No root cause determine due complaint is not attributable to the manufacturing process.A notification was sent to the supplier.A dhr review was not completed as the lot number was not provided.
|
| |
|