MAUDE Adverse Event Report: OAKDALE LEVEL 1 LOW FLOW HOTLINE DISPOSABLE INJECTION SET; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OAKDALE LEVEL 1 LOW FLOW HOTLINE DISPOSABLE INJECTION SET; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Catalog Number L-70NI
Medical Device Problem Codes Break (1069); Material Separation (1562); Connection Problem (2900)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/01/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that due to damage to the lure portion of the connection, it could not be connected to the route and its use was discontinued.No patient injury or clinical affects was reported.
 
Additional Manufacturer Narrative
Other text: b3: month and year of event have been provided, day is unknown.D4: lot number, expiration date, and h4: device manufacture date is unknown, no information has been provided to date.D3, g1,g2 email is:(b)(6) investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Additional Manufacturer Narrative
Other, other text: device evaluation: one sample was received in used condition.A visual inspection found a crack in the female luer connector.During the functional testing, a leak test was done and during the third attempt the female luer presented damage in the connection part.However, the damage was not similar to the sample reported.Based on the replication of the failure mode results, the crack reported is not attributable to the manufacturing process.No root cause determine due complaint is not attributable to the manufacturing process.A notification was sent to the supplier.A dhr review was not completed as the lot number was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 LOW FLOW HOTLINE DISPOSABLE INJECTION SET
Common Device Name
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
reed covert
MDR Report Key18241245
Report Number3012307300-2023-10887
Device Sequence Number3225248
Product Code LGZ
UDI-Device Identifier30695085407502
UDI-Public30695085407502
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K911383
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberL-70NI
Was Device Available for Evaluation? No
Date Returned to Manufacturer11/16/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/06/2023
Supplement Date Received by Manufacturer11/06/2023
Initial Report FDA Received Date11/30/2023
Supplement Report FDA Received Date08/27/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-