| Model Number |
M0046DYNTP0010 |
| Medical Device Problem Codes |
Material Deformation (2976); Material Integrity Problem (2978); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/06/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial reporter phone: (b)(6).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the tip of the device was fractured.During a radiofrequency ablation procedure, a dynamic tip electrode catheter was selected for use.It was reported that the junction of the catheter was kinked.Also, when unpacking the device, it was noticed the tip of the catheter was fractured.Device was replaced and the issue was resolved.Procedure was able to be completed successfully without any patient complications.Catheter is expected to be returned for further analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Dynamic tip electrode catheter was evaluated by boston scientific.Analysis of returned product revealed that the unit showed no evidence or defects that could have contributed to the tip damaged-kinked event; consequently, not confirming the reported clinical observation of tip fracture.However, during the product analysis it was observed that the plunger of the device was broken.The cause of the issue seen in the field could not be determined.Initial reporter phone: (b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that the tip of the device was fractured.During a radiofrequency ablation procedure, a dynamic tip electrode catheter was selected for use.It was reported that the junction of the catheter was kinked.Also, when unpacking the device, it was noticed the tip of the catheter was fractured.Device was replaced and the issue was resolved.Procedure was able to be completed successfully without any patient complications.Catheter was returned to boston scientific for laboratory analysis.
|
| |
|
Event or Problem Description
|
|
It was reported that the tip of the device was fractured.During a radiofrequency ablation procedure, a dynamic tip electrode catheter was selected for use.It was reported that the junction of the catheter was kinked.Also, when unpacking the device, it was noticed the tip of the catheter was fractured.Device was replaced and the issue was resolved.Procedure was able to be completed successfully without any patient complications.Catheter was returned to boston scientific for laboratory analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
"**udi related data quality updates only** this report is being filed as a correction in response to an fda request for the complete the unique identifier (udi) #." dynamic tip electrode catheter was evaluated by boston scientific.Analysis of returned product revealed that the unit showed no evidence or defects that could have contributed to the tip damaged-kinked event; consequently, not confirming the reported clinical observation of tip fracture.However, during the product analysis it was observed that the plunger of the device was broken.The cause of the issue seen in the field could not be determined.Initial reporter phone: (b)(6).
|
| |
|