| Catalog Number |
SVMB13 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/28/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported the implant fixture at tooth site #13, was noted to be fractured at platform level.Therefore, the implant was removed.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint number cmp- (b)(4).A4: patient weight unknown / not provided g4: additional pma/510(k) number ¿ k011028|k013227 product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number (b)(4).Zimvie received one (1) svmb13, (impl scr-v mini sbm 3.3mm 3.5mm 13mm) for evaluation.Visual evaluation of the as returned device identified the implant with signs of use, apparent bone debris on external threads.Implant fractured at the collar.A fractured screw was identified inside the implant.Functional testing to recreate the reported event could not be performed due to the nature of the device & event.This complaint refers to the specific device being investigated for this complaint record.Dhr review was completed for the subject lot number 63731201.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 63731201 for similar events and no other complaint was identified.Review completed utilizing keywords: fracture implant the customer did not submit images for the reported event.Ifu review: documents reviewed: instructions for use for screw-vent® implants - 9665 rev.0-09/14.Information identified: warnings contraindications precautions breakage based on the investigation and risk management file review, the most likely root causes determined from the investigation are clinician error, patient factors.Therefore, based on the available information, a device malfunction did occur.The reported event was confirmed with all the available information.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.Investigation results.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number (b)(4).The following fields have been updated: b4: date of this report.B5: describe event or problem.D1: brand name updated.D3: manufacturer email address.D9: device availability.G1: contact office (and manufacturing site for devices) contact name and email.G3: date received by manufacturer.G6: type of report.H1: type of reportable event.H2: follow up type.H3: device evaluated by manufacturer.H4: manufacturing date updated.H10: additional narrative.This report is submitted to update manufacturing date h3 other text : investigation results.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Search Alerts/Recalls
|