• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NALU MEDICAL INC NALU NEUROSTIMULATION SYSTEM; PERIPHERAL NERVE STIMULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NALU MEDICAL INC NALU NEUROSTIMULATION SYSTEM; PERIPHERAL NERVE STIMULATOR Back to Search Results
Model Number 71019
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Inadequate Pain Relief (2388)
Date of Event 11/01/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of records found no previous communications or complaints from this patient related to the nalu system.The patient was offered troubleshooting prior to surgical intervention and patient declined.At the explant there was no visual indication of any device or component malfunction or migration, however there was no imaging or other testing performed prior to explant in order to confirm.The device was damaged during the explant procedure and thus is not in a condition that is conducive to testing by the firm.
 
Event or Problem Description
Patient was implanted with the nalu peripheral nerve stimulator (pns) system on (b)(6) 2022 targeting the suprascapular nerve to treat shoulder pain.On (b)(6) 2023 the firm was notified by the surgical center that the patient was scheduled for a full system explant.Patient expressed that they did not feel like adequate pain relief was being achieved, patient wants to explant the pns system and pursue a spinal cord stimulator system.No further information was shared with the firm.A full system explant took place on (b)(6) 2023.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NALU NEUROSTIMULATION SYSTEM
Common Device Name
PERIPHERAL NERVE STIMULATOR
Manufacturer (Section D)
NALU MEDICAL INC
2320 faraday avenue
suite 100
calsbad CA 92008
Manufacturer (Section G)
NALU MEDICAL, INC
2320 faraday ave
suite 100
carlsbad
Manufacturer Contact
terry villarba
2320 faraday avenue
suite 100
carlsbad, CA 92008-7241
7604482360
MDR Report Key18243746
Report Number3015425075-2023-00295
Device Sequence Number6582450
Product Code GZF
UDI-Device Identifier00812537035604
UDI-Public01008125370356041122053117250531
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K183579
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2025
Device Model Number71019
Device Catalogue Number71019
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/01/2023
Initial Report FDA Received Date11/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/31/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-