| Catalog Number |
11895354160 |
| Medical Device Problem Codes |
False Negative Result (1225); Device Difficult to Setup or Prepare (1487)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/03/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The meter and test strips were requested for investigation.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
The initial reporter received questionable leukocyte results for four urine patient samples tested on the urisys 1100 urine analyzer us with serial number (b)(6).The reporter stated that the meter results were not aligned with the patients' symptoms prompting them to send the urine samples to the laboratory.Results for the four urine patient samples: the meter results were negative for leucocytes.The laboratory results were positive.The actual laboratory results were requested but not provided.The patients were treated based on the laboratory results.
|
| |
|
Additional Manufacturer Narrative
|
|
The meter and test strips were not received for investigation.The retention material of the lot was tested; no false negative results were obtained.The retention material of the lot fulfilled the requirements.The investigation did not identify a product problem.
|
| |
|