Model Number FV701T
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 11/04/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a prosa (#fv701t) was implanted combined with a progav 2.0 (#fx410t) and a shuntassistant (fv252t) during a procedure performed on (b)(6) 2020.According to the complainant, the valve caused an underdrainage.The patient underwent a revision procedure.The complainant device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 34 years; height: 160 cm; weight: 65 kg; gender: female.Same event as medwatch#3004721439-2023-00355 and #3004721439-2023-00357.
 
Additional Manufacturer Narrative
Investigation: visual inspection: during the investigation, a deformation of the outer housing could be determined during the visual inspection.The measurement of the plane parallelism has confirmed this with a value of -0.054 mm - outside the tolerance of 0 ± 0.02 mm.Permeability test: a permeability test has shown that all components are permeable.Computer control test: to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer control testing bench which simulates a cerebrospinal fluid flow.The valve is tested in both the horizontal as well as the vertical position.Adjustment test: the prosa valve was tested and is adjustable to all specified pressures.Braking force and brake function test: the brake functionality test has shown that the brake function is fully operational and the braking force is within the given tolerances.Internal inspection : after dismantling, no deposits were found in the prosa.Results: based on our investigation results, we cannot determine a functional deviation or other abnormalities.The cause of the aforementioned functional impairment is not known to us at the time of the examination.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROSA VALVE
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key18247000
Report Number3004721439-2023-00356
Device Sequence Number7191320
Product Code JXG
UDI-Device Identifier04041906133023
UDI-Public4041906133023
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K161853
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/13/2025
Device Model NumberFV701T
Device Catalogue NumberFV701T
Device Lot Number20047837
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/09/2023
Initial Report FDA Received Date12/01/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/15/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age34 YR
Patient SexFemale
Patient Weight65 KG
-
-