Investigation: visual inspection: during the investigation, a deformation of the outer housing could be determined during the visual inspection.The measurement of the plane parallelism has confirmed this with a value of -0.054 mm - outside the tolerance of 0 ± 0.02 mm.Permeability test: a permeability test has shown that all components are permeable.Computer control test: to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer control testing bench which simulates a cerebrospinal fluid flow.The valve is tested in both the horizontal as well as the vertical position.Adjustment test: the prosa valve was tested and is adjustable to all specified pressures.Braking force and brake function test: the brake functionality test has shown that the brake function is fully operational and the braking force is within the given tolerances.Internal inspection : after dismantling, no deposits were found in the prosa.Results: based on our investigation results, we cannot determine a functional deviation or other abnormalities.The cause of the aforementioned functional impairment is not known to us at the time of the examination.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
|