| Model Number |
A22251C |
| Medical Device Problem Codes |
Break (1069); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
11/01/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A company representative, on behalf of the customer, reported to olympus that during a transurethral resection of the prostate, the monopolar rollerball of the hf-resection electrode broke off in patient.Despite multiple attempts to obtain additional information regarding the device and event, none was obtained.
|
| |
|
Additional Manufacturer Narrative
|
|
D1, full name: hf-resection electrode, roller, 24-28 fr., 12° and 30°, sterile, single use, 12 pcs.This report was submitted by the importer under the importer's report number: 2429304-2023-00376 the investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information, based on the legal manufacturer's final investigation.A review of the device history record found no deviations, that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely, that the reported event occurred, due to wear and tear.Additionally, the loop at the distal end of the electrode wears out, during use and may break, burn or melt.However, the subject device was not returned.And the root cause of the reported event could not be determined.Olympus will continue to monitor field performance for this device.
|
| |
|