• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CHRISTOPH MIETHKE GMBH & CO KG PAEDIGAV SYSTEM W/PRECHAMB.ON VALVE 4/19; HYDROCEPHALUS MANAGEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CHRISTOPH MIETHKE GMBH & CO KG PAEDIGAV SYSTEM W/PRECHAMB.ON VALVE 4/19; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FV303T
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 11/08/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a paedigav (#fv303t) was implanted together with an progav 2.0 during a procedure performed in (b)(6) 2021.According to the complainant, the valve caused an underdrainage.The patient underwent a revision procedure.The complainant device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 68 years.Gender: female.Same event/patient as medwatch#3004721439-2023-00367.
 
Additional Manufacturer Narrative
Investigation: visual inspection: during the investigation, no significant deformation or damage of the valve was determined.Permeability test: a permeability test has shown that the valve is permeable.Computer control test: to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer control test bench which simulates a cerebrospinal fluid flow.The valve is tested in both the horizontal as well as the vertical positions.The results show that the valve is operating within the accepted tolerance in the horizontal but not in the vertical position.An accelerated outflow of the paedigav could be determined.Internal inspection : after dismantling of the valve, visible deposits were found.To make the proteins / deposits in the valve more visible, they were colored using a staining solution.Result: based on our investigation results, we can determine an accelerated outflow in the valve.The determined deposits can be named as the cause for the accelerated outflow.Organic material in the cerebrospinal fluid can influence the function temporarily and are known side effects in hydrocephalus therapy.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PAEDIGAV SYSTEM W/PRECHAMB.ON VALVE 4/19
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key18249700
Report Number3004721439-2023-00360
Device Sequence Number6645020
Product Code JXG
UDI-Device Identifier04041906131180
UDI-Public4041906131180
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K011030
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/27/2024
Device Model NumberFV303T
Device Catalogue NumberFV303T
Device Lot Number20042172
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/21/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/09/2023
Initial Report FDA Received Date12/01/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/28/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
Patient SexFemale
-
-