• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH SILIC HAND W/AO/ASIF-QC; FORCEPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH SILIC HAND W/AO/ASIF-QC; FORCEPS Back to Search Results
Catalog Number 03.118.111
Medical Device Problem Codes Device Slipped (1584); Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/10/2023
Type of Reportable Event Malfunction
Event or Problem Description
Device report from synthes reports an event in colombia as follows: it was reported on (b)(6) 2023, that during surgery, the specialist drills to pass a cortical screw but refers that the handle has no force and it is loose apparently, and the tip does not hold the screws causing difficulty inserting them into the plate and also causes discomfort in the specialist, it was delayed approximately 5 minutes of the procedure, the surgery was successful.It was partially resolved using the handle of ref 01.118.226 corresponding to the instrumental of 2.7.There was no way to resolve the issue with the tip.There was no harm to the patient.This report is for one (1) silic hand w/ao/asif-qc this is report 1 of 1 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9: complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: part number: 03.118.111, lot number: t195268, manufacturing site: tuttlingen, release to warehouse date: 30-jul-2020.A review of the device history records was performed for the finished device lot number, and no non-conformances were identified.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that silic hand w/ao/asif-q presented signs of normal use.The knob of the device is loose and presents a creaking sound.A functional test was performed with mating device 314.116 scrdriver shaft 3.5 t15 self-holding f/a revealed that although the devices assemble, they do not have a tight fit.Unable to assemble condition was not able to be replicated.A dimensional inspection was not performed for the silic hand w/ao/asif-q since it was not applicable.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the silic hand w/ao/asif-q would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SILIC HAND W/AO/ASIF-QC
Common Device Name
FORCEPS
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES TUTTLINGEN
unter hasslen 5
tuttlingen CO 78532
GM   78532
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key18250834
Report Number8030965-2023-15217
Device Sequence Number17792314
Product Code HTD
UDI-Device Identifier07611819521407
UDI-Public(01)07611819521407
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.118.111
Device Lot NumberT195268
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/10/2023
Supplement Date Received by Manufacturer01/03/2024
Initial Report FDA Received Date12/01/2023
Supplement Report FDA Received Date01/19/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/30/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SCRDRIVER SHAFT 3.5 T15 SELF-HOLDING F/A; UNK - PLATES: TRAUMA; UNK - SCREWS: CORTICAL : TRAUMA
Patient SexFemale
-
-