• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS AG DAA DOUBLE OFFSET ADAPTER RIGHT 80/45; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS AG DAA DOUBLE OFFSET ADAPTER RIGHT 80/45; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Catalog Number 75102240
Medical Device Problem Codes Mechanics Altered (2984); Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/15/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, when attempting to explant an undersized trial broach during a thr surgery, the locking mechanism of a daa double offset adapter right 80/45 was not optimal.It was pushing the broach further down the femur.The procedure was resumed after a significant delay, using a combination of bone hook, osteotome, ronguer and cobb.No injury was reported as a consequence of this issue.
 
Additional Manufacturer Narrative
Internal complaint number: case (b)(4).
 
Additional Manufacturer Narrative
It was reported that, when attempting to explant an undersized trial broach during a total hip replacement surgery, the locking mechanism of a daa double offset adapter right 80/45 was not optimal.It was pushing the broach further down the femur.No injury was reported as a consequence of this issue.The device was not returned for investigation.A product evaluation was not possible.No batch number was communicated so the production history review was not possible.Due to an unknown batch number, the review of historical complaints was performed on product number basis only, revealing 1 additional complaints over the past 12 months with similar failure mode.Due to insufficient information it is not possible to perform a review of past corrective actions.A review of the risk management documentation verifies the failure mode, occurrence and severity of the reported issue.The performed investigation does not lead to an accurately determined cause.Based on the available information it is not possible to investigate whether the reported device met manufacturing specifications upon release for distribution.Due to insufficient information, it is not possible to indicate factors which could have contributed to the reported event.According to document "processing (cleaning, disinfection and sterilization) of instruments from smith & nephew orthopaedics ag" (lit.N°03389-en 1363 v3 11/19), all devices must be inspected and controlled for proper functioning after cleaning/disinfection.There is no need for further actions because of the limited information provided.Nevertheless, smith + nephew will continue to monitor this device for similar issues.This complaint will be reopened should additional information or the device be received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DAA DOUBLE OFFSET ADAPTER RIGHT 80/45
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ  CH-5000
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ   CH-5000
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18252186
Report Number9613369-2023-00222
Device Sequence Number12539854
Product Code JDI
UDI-Device Identifier00885556535394
UDI-Public00885556535394
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K123598
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number75102240
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/15/2023
Supplement Date Received by Manufacturer12/19/2023
Initial Report FDA Received Date12/01/2023
Supplement Report FDA Received Date12/20/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Unknown
Patient Sequence Number1
Patient SexMale
Patient Weight172 KG
-
-