| Catalog Number |
362420 |
| Medical Device Problem Code |
Device Alarm System (1012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/07/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
|
| |
|
Event or Problem Description
|
|
As reported by the user facility: brief inquiry description: infusomat giving ail alarms during infusion.Detailed inquiry description: infusomat giving ail alarms during infusion.Numerous sets triggering ail alarms on pumps when no air bubbles are present.No injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).No sample was provided for evaluation.Due to complaints of this nature, corrective actions were performed to revise specifications which limit unit-to-unit variation in tubing outer diameters and increased the length of the tubing that contacts the air sensor.The reduced variation in tubing outer diameter and length revision improves the coupling between the administration set and the air-in-line sensor reducing the potential for false air-in-line alarms.B.Braun medical inc.Has initiated a voluntary medical device correction 2523676-9/26/23-004-c for multiple batches of infusomat® pump sets manufactured between january 1, 2022, and august 17, 2023.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|
| |
|
Search Alerts/Recalls
|